Amyloid-beta 42 adsorption following serial tube transfer.

Toombs J, Paterson RW, Schott JM, Zetterberg H.
Journal   Alzheimers Res Ther.
Species  
Analytes Measured  
Matrix Tested   Cerebrospinal fluid (CSF)
Year   2014
Volume   6
Page Numbers   5
Application  
Abstract
INTRODUCTION: Cerebrospinal fluid (CSF) amyloid-beta 38 (Aβ38), 40 (Aβ40), 42 (Aβ42) and total tau (T-tau) are finding increasing utility as biomarkers of Alzheimer's disease (AD). The purpose of this study was to determine whether measured CSF biomarker concentrations were affected by transfer of CSF between tubes, and whether addition of a non-ionic surfactant mitigates any observed effects.

METHODS: AD and control CSF was transferred consecutively between polypropylene tubes. Aβ peptides and T-tau were measured with and without addition of Tween 20 (0.05%).

RESULTS: Measured concentrations of Aβ42 decreased by approximately 25% with each consecutive transfer. Measured concentrations of Aβ38 and Aβ40 were also observed to decrease significantly with each consecutive transfer (approximately 16% loss per transfer). Measured concentrations of T-tau also decreased significantly, but at much smaller magnitude than the Aβ peptides (approximately 4% loss per transfer). The addition of Tween 20 mitigated this effect in all samples.

CONCLUSIONS: Consecutive CSF transfer between tubes has a significant impact on the measured concentration of all Aβ peptides, and significant effect of lesser magnitude on T-tau. This would be sufficient to alter biomarker ratios enough to mislead diagnosis. The introduction of Tween 20 at the initial aliquoting stage was observed to significantly mitigate this effect.

View Publications

Related Products

V-PLEX Human Aβ42 Peptide (6E10) Kit
Aβ42 (6E10) | Human
Singleplex
V-PLEX Plus Human Aβ42 Peptide (6E10) Kit
Aβ42 (6E10) | Human
Singleplex
V-PLEX Plus Aβ Peptide Panel 1 (6E10) Kit
Aβ38 (6E10), Aβ40 (6E10), Aβ42 (6E10) | Human
Multiplex
V-PLEX Aβ Peptide Panel 1 (6E10) Kit
Aβ38 (6E10), Aβ40 (6E10), Aβ42 (6E10) | Human
Multiplex
Browse Our Products

By Analytes
By Applications
Search
Meso Scale Japan 株式会社