Novel Multiplexed Serology Assays for Detection of IgG Antibodies against Monkeypox (mpox) and Vaccinia Viruses

Download ↓

1. Abstract
Monkeypox (mpox) virus (MPXV) spreads through skin-to-skin contact, causing painful lesions. The 2022 outbreak resulted in the Congo’s worst surge on record, with the subsequent world-wide spread causing nearly 20,000 suspected cases and 820 suspected deaths. Effective tools for understanding MPXV immune responses are needed for developing timely and effective MPXV-specific vaccines, as well as for understanding the immune correlates of protection from natural infection and/or vaccination. Despite a rapid increase in the number and availability of serology assays that can detect antibodies against MPXV, there is limited information available on their performance and validation status. In addition, most of these assays are low throughput and measure responses to a single antigen, which cannot capture the breadth of antibody responses to MPXV. Here we present a validated, quantitative, multiplexed serology assay to measure antibody responses towards 5 MPXV and 5 vaccinia virus (VACV) variant recombinant proteins. Vaccinia is included in the panel due to its common ancestry to MPXV and the prevalence of smallpox vaccination, which is expected to prevent or reduce the severity of MPXV infection. Viral antigens that elicit strong T cell and B cell immune responses were chosen for the panel and include the receptor binding site for MPXV (A29L) and VACV (A27L), outer envelope proteins (MPXV: B6R, A35R and VACV: B5R, A33R) and inner membrane proteins (MPXV: M1R, E8L and VACV: L1R, D8L). The assay uses a 10-spot 96-well plate coated with the 5 MPXV and 5 VACV antigens, along with an electrochemiluminescent (ECL) detection system. The assay simultaneously detects IgG antibodies to all 10 proteins. Specificity was assessed using purchased serum sample sets that were either collected pre-epidemic from aged smallpox-vaccinated individuals, or during the outbreak from convalescent individuals who recently recovered from MPXV infection. The multiplex MPXV ECL serology assay allows for sensitive, high throughput, and simultaneous measurement of IgG levels to multiple antigens in human sera, supporting its use in research, epidemiology, and vaccine testing.

View More View Less

2. Methods

MSD Technology
MSD’s electrochemiluminescence detection technology uses SULFO-TAG labels that emit light upon electrochemical stimulation initiated at the electrode surfaces of MULTI-ARRAY and MULTI-SPOT microplates.

Electrochemiluminescence Technology

Electrochemiluminescence Technology

  • Minimal non-specific background and strong responses to analyte yield high signal-to-background ratios.
  • The stimulation mechanism (electricity) is decoupled from the response (light signal), minimizing matrix interference.
  • Only labels bound near the electrode surface are excited, enabling non-washed assays.
  • Labels are stable, non-radioactive, and directly conjugated to biological molecules.
  • Emission at ~620 nm eliminates problems with color quenching.
  • Multiple rounds of label excitation and emission enhance light levels and improve sensitivity.
  • Carbon electrode surface has 10X greater binding capacity than polystyrene wells.
  • Surface coatings can be customized.

The V-PLEX Orthopoxvirus Panel 1 (IgG) Kit includes a multiplex panel to detect antibodies to antigens from mpox and vaccinia viruses, including variants related to viral structures such as the receptor binding region, inner membrane and outer membrane regions of the viral core. Plates are provided with antigens on spots in the wells of a 96-well plate. Antibodies in the sample bind to the antigens on the spots and anti-human antibodies (IgG) conjugated with MSD SULFO-TAG are used for detection. The plate is read on an MSD instrument, which measures the light emitted from the MSD SULFO-TAG.

Orthopoxvirus Panel 1 (IgG) Kit

The kit includes a calibrator for quantitation, controls, plate(s), detection antibody (anti-human IgG), and all other reagents necessary to conduct the assay. Plates are coated with 5 mpox antigens (M1R, E8L, B6R, A35R, and A29L) and five analogous vaccinia viral antigens (A27L, A33R, B5R, D8L, and L1R). Including both mpox and vaccinia ortholog antigens in a single panel allows researchers to evaluate infection/vaccination status and potential cross-protection.

Plates are coated with 5 mpox antigens (M1R, E8L, B6R, A35R, and A29L) and five analogous vaccinia viral antigens (A27L, A33R, B5R, D8L, and L1R).

3. Results

a. Signal Reproducibility
Orthopoxvirus plates tested for signal reproducibility across wells and plates were analyzed on a per-batch basis and passed a 20% CV intraplate average specification for all ten multiplexed assays. Representative data demonstrated consistent performance using a mid-level calibration standard (Cal-03) in a standard assay protocol.

Orthopoxvirus plates tested for signal reproducibility across wells and plates

b. Specificity and Sample Panel
Screening was performed to identify potential antibodies capable of distinguishing between mpox antigens, to be used to evaluate plate coating specificity. Most antibodies recognized their mpox target as well as the target’s vaccinia ortholog. Selected antibodies (monoclonals and polyclonals) were used to build specificity reagents to confirm plates were coated correctly, based on a specification of ≤ 1% ECL signal of nonspecific binding (NSB). Each antigen’s identity was confirmed prior to coating via sequencing and other analytical methods.

Screening was performed to identify potential antibodies capable of distinguishing between mpox antigens, to be used to evaluate plate coating specificity.

To further distinguish between antigens, we also screened commercially-sourced human serum samples that were collected pre-epidemic (smallpox antibody positive, smallpox/mpox negative) and convalescent for mpox. Results calculated from two plate lots and highlighted in green showed at least 20% difference in ECL signal between ortholog pairs.

we also screened commercially-sourced human serum samples that were collected pre-epidemic (smallpox antibody positive, smallpox/mpox negative) and convalescent for mpox.

c. V-PLEX Orthopoxvirus Panel 1 Representative Data (Calibrator, Controls and Sample Testing)
Human serum was used to build Orthopoxvirus Serology Calibrator 1 and Orthopoxvirus Serology Controls 1, 2, and 3 for use in the V-PLEX Orthopoxvirus Serology Kit. Selectivity was assessed using a set of purchased sera (as described in Section b), to confirm reproducible signal measurements for calibrator, controls and samples.
Using MSD assigned concentrations for the Orthopoxvirus Serology Calibrator 1, signal concentration curves for two lots confirmed that lot-to-lot variability was <20% CV. Functional IgG measurements were performed for each assay, and Hill slopes for the lots were tracked closely.
Each run consisted of calibrator curves, controls, Upper Limit of Quantification (ULOQ) samples, and Lower Limit of Quantification (LLOQ) samples (sample panel, serum-01-12 from Section b).

d. V-PLEX Orthopoxvirus Panel 1 Representative Data (Calibrator, Controls and Sample Testing) con’t
Representative calibrator curves show a dynamic range of 3-4 logs for all assays. Orthopoxvirus Panel 1 MPXV and VACV antigens demonstrated good precision with specification of ≤ 20%.

Orthopoxvirus Panel 1 MPXV and VACV antigens demonstrated good precision with specification of ≤ 20%.

Calculated percent recovery using each control’s concentration was within specification as indicated below. The precision met the specification of ≤ 20%, indicating control recoveries were very similar across two plate lots.
Orthopoxvirus Serology Control 3 is a “negative” control with concentration values <0.1 AU/mL (Arbitrary Units/mL) and are less than LLOQ (data not shown).

Orthopoxvirus Serology Control 3 is a “negative” control with concentration values <0.1 AU/mL (Arbitrary Units/mL) and are less than LLOQ (data not shown).

IgG functional testing using a set of human serum samples (as described in Sections b and c) demonstrated precision that met specification (≤ 20%).
MPXV and VACV specific IgG were measured at higher concentrations in all human serum samples compared to the negatives which have concentrations ~0.01 – 0.10 AU/mL.

MPXV and VACV specific IgG were measured at higher concentrations in all human serum samples compared to the negatives which have concentrations ~0.01 – 0.10 AU/mL.

4. Conclusions

Mpox infection has been a global public health concern since the multi-country outbreak in 2022. Vaccination against smallpox was protective; however, due to the global cessation of smallpox vaccination campaigns following the disease's eradication, certain populations may now be more susceptible to mpox. Studies aimed at better understanding the epidemiology, sources of infection, and transmission patterns in regards to mpox would benefit from a multiplex serological method for researching at-risk populations. This method would help distinguish between vaccinated and/or mpox-infected samples versus negative samples, and provide researchers with an alternative to PCR.
MSD was already positioned to provide tools for COVID outbreaks, and with the launch of the V-PLEX Orthopoxvirus Panel 1 in 2023, MSD was able to offer the first multiplex serology kit for mpox research applications. This multiplex panel is based on mpox and vaccinia viral antigens coated on MSD 96-well MULTI-SPOT plates. The kit includes a calibrator for quantitation (Orthopoxvirus Serology Calibrator 1), controls (Orthopoxvirus Serology Control Pack), antigen-coated plate(s), detection antibody (SULFO-TAG Anti-Human IgG), and all other reagents necessary to conduct a quantitative research use only (RUO) assay. The kit has various applications, including correlation of protection studies for vaccine development, study of long-term humoral immunity to mpox infection following natural infection or vaccination, and orthopoxvirus vaccination or surveillance (Hicks, et al., 2024; Macedo Cincotta, et al., 2024).
The V-PLEX Orthopoxvirus Serology Kit is a validated, quantitative, multiplexed RUO serology kit that accurately and precisely quantifies IgG antibodies against the MPXV and VACV proteins (VACV A27L, VACV A33R, VACV B5R, VACV D8L, VACV L1R, MPXV M1R, MPXV E8L, MPXV B6R, MPXV A35R and MPXV A29L).